| 联影ARMS技术对盆腔磁共振扫描图像质量的影响 |
| Impact of combined imaging ARMS technology on the image quality of pelvic magnetic resonance imaging |
| 投稿时间:2026-01-28 修订日期:2026-10-21 |
| DOI: |
| 中文关键词: 磁共振成像 ARMS技术 盆腔 图像质量 |
| 英文关键词: Magnetic resonance imaging ARMS technology Pelvic cavity Image quality |
| 基金项目:无 |
|
| 摘要点击次数: 78 |
| 全文下载次数: 0 |
| 中文摘要: |
| 【】 目的 分析联影ARMS技术对盆腔磁共振(MRI)扫描图像质量的影响。方法 选取2024年2月~2025年2月于我院接受盆腔MRI扫描的女性患者116例为研究对象,均行联影MRI常规序列和ARMS序列扫描,扫描结束后由2名放射科医师以盲法阅片,采用Likert-type量表5级评分法对图像进行主观质量评价,采用信噪比(SNR)和对比度噪声比(CNR)、均匀度、扫描时间、空间分辨率对图像进行客观质量评价。结果 经医师1评价,ARMS序列伪影抑制、解剖结构清晰度及图像整体质量评分均高于MRI常规序列,差异有统计学意义(P<0.05)。经医师2评价,ARMS序列伪影抑制、解剖结构清晰度及图像整体质量评分均高于MRI常规序列,差异有统计学意义(P<0.05)。2名医师的评价结果具有高度一致性(Kappa值>0.80)。ARMS序列SNR、CNR、均匀度均高于MRI常规序列,扫描时间短于MRI常规序列,差异有统计学意义(P<0.05)。ARMS序列空间分辨率高于MRI常规序列,差异有统计学意义(P<0.05)。结论 联影ARMS技术能有效提高盆腔MRI图像质量,为临床诊断提供更精确的影像学依据。 |
| 英文摘要: |
| Objective To analyze the impact of combined imaging acquisition and reconstruction with motion suppression (ARMS) technology on the image quality of pelvic magnetic resonance imaging (MRI). Methods A total of 116 female patients who received pelvic MRI examination in the hospital from February 2024 to February 2025 were selected as the research subjects. Both conventional MRI sequence scan and ARMS sequence scan were performed. Two radiologists read the images using the blind method. The 5-level Likert-type scale was used for subjective quality assessment of images. The objective quality of the image was evaluated by using the signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), uniformity, scanning time and spatial resolution. Results According to evaluation by radiologist one, the scores for artifact suppression, clarity of anatomical structures, and overall image quality of ARMS sequences were significantly higher than those of conventional MRI sequences (P<0.05). According to evaluation by physician two, the scores for artifact suppression, clarity of anatomical structures, and overall image quality of ARMS sequences were significantly higher than those of conventional MRI sequences (P<0.05). The evaluation results of the two physicians were highly consistent (Kappa value >0.80). The SNR, CNR and uniformity of the ARMS sequence were all higher than those of the conventional MRI sequence, and the scanning time was shorter than that of the conventional MRI sequence (P<0.05). The spatial resolution of the ARMS sequence was higher than that of the conventional MRI sequence (P<0.05). Conclusion The combined imaging ARMS technology can effectively improve the quality of pelvic MRI images and provide more accurate imaging evidence for clinical diagnosis. |
|
View Fulltext
查看/发表评论 下载PDF阅读器 |
| 关闭 |
|
|
|